Ventilator,medical,adult-child,w/access
.png)
Product Code : SCL-MAI-12220
GENERAL DESCRIPTION
Ventilator, Intensive Care, for adult, paediatric and neonatal patients, AC and battery powered, with accessories
INTENDED USE
Mechanical ventilators are life support devices that move gas (e.g., air and/or oxygen) to and from a patient's lungs. These devices may provide temporary or permanent respiration for patients who cannot breathe on their own, or who require assistance maintaining adequate ventilation because of illness, trauma, congenital defects, or the effects of drugs (e.g., anaesthetics). In most cases, mechanical ventilators are used for a short period of time (a few days to a few weeks) to deliver pressurized medical gases to the patient's lungs to support gas exchange and rest ventilatory muscles until the patient is able to breathe without mechanical assistance. Some patients, however, require permanent ventilatory support.
TECHNICAL SPECIFICATIONS
Automatic ventilator for adult, paediatric and neonatal patients.
Equipped with an internal turbine to eliminating the need for external supply of compressed air.
The unit should be mounted on a sturdy trolley with handles.
The trolley is equipped with 4 antistatic swivel castors.
At least two castors have been equipped with brakes.
Designed for frequent and easy dismount and disinfection with hospital-grade products.
Adjustable patient-circuit support arm.
Equipped with integrated oximeter for FiO?.
Equipped with an Air/O? mixer.
Ability to calculate intrinsic PEEP volume, occlusion pressure (calculated manually) and inflection points.
Unit is fitted with temperature sensor.
Unit is fitted with patient circuit.
Unit is fitted with condensate collection device.
Unit is fitted with anti-bacterial filter.
Automatic compliance and leakage compensation for circuits and tubes.
Equipped with an autoclavable expiration block.
Reusable, removable and disinfectable expiration valve/flow sensor.
The unit accepts inlet gas supply pressures between 2.8 bar to 6 bar (40 to 87 psi).
Gas inlet connections are standard DISS (Diameter Index Safety Standard)
connectors.
The unit is equipped with a servo-controlled humidifier.
The humidifier has a heat exchanger which allows for temperatures between 28°C and 39°C at the Y-piece.
The humidifier temperature monitoring has an accuracy of at least ± 1°C.
The humidifier has the ability to switch between passive and active humidification.
In active mode the humidity levels the humidifier can provide are in the range of 33 mg H?O/L to 44 mgH?O/L.
In passive mode the humidity levels the humidifier can provide are at least 30 mg H?O/L.
The resistance of the humidifier is no less than 0.5 cm H?O.
The compliance of the humidifier is below 1.1 mL/cm H?O.
The humidifier has a sterilisable humidifier chamber.
Equipped with a display (colour, flat panel ≥ 10 Inch) showing:
- Operational status.
- Set and measured parameters and values.
- Waveforms.
- Alarms/errors such as ventilation parameters, power, and system events, etc.
Equipped with a drugs nebulizer.
The unit is equipped with self-diagnosis and sensor calibration processes.
Built-in rechargeable battery, autonomy > 1 hour (standard ventilation, excluding compressor).
Automatic switch from mains to battery in case of power failure.
Automatic battery charge when mains connection is re-established.
Power requirements: units should be available for 100 and 240 V - 60 / 50 Hz.
VENTILATION MODES
The unit supports the following ventilation modes:
- Assist Control mode.
- Pressure Controlled Ventilation (PCV), pressure-controlled breaths.
- Volume Controlled Ventilation (VCV), volume-controlled breaths.
- Pressure-Regulated Volume Control (PRVC).
- Continuous Mandatory Ventilation (CMV).
- Synchronised Intermittent Mandatory Ventilation (SIMV), volume-controlled breaths, pressure-controlled breaths, and pressure support.
- Continuous Positive Airway Pressure mode (CPAP) and pressure support.
- Apnoea-backup ventilation mode.
- Spontaneous.
- (Volume targeted) Pressure controlled ventilation (APCV-TV).
- (Volume targeted) Pressure support ventilation (PSV-TV).
- Pressure Support Ventilation (PSV).
CONTROLS AND SETTINGS
The unit supports the following controls and settings:
- Positive End Expiration Pressure (PEEP): 0 - 50 cmH?O.
- Pressure Support: 0 - 80 cmH?O.
- Peak Inspiration Pressure: 2 - 80 cmH?O.
- Tidal Volume: 2 - 3,000 ml.
- Respiration rate: 4 - 150 bpm.
- SIMV Respiration rate: 1 to 60 bpm.
- I/E ratio 1:10 to 4:1.
- Inspiratory Flow: 1 - 190 l/min
- Inspiratory Time: 0.036 – 9.6 sec.
- FiO?: 21 - 100%.
- Inspiration pause time: 0 – 60% of the inspiration time.
- Inspiration rise time adjustable in four steps.
- Expiration time: 0.08 – 10.9 sec.
- Apnoea time adjustable: 5 to 60 Sec.
MONITORED AND MEASURED PARAMETERS
The below listed parameters are monitored and measured:
- I/E ratio.
- Inspiratory and expiratory times.
- Tidal Volume.
- Inspiratory and expiratory volumes.
- Minute Volume (spontaneous and mechanical).
- Respiratory rate (spontaneous and mechanical).
- Total frequency.
- Mean airway pressure.
- Peak inspiratory pressure.
- PEEP.
- FiO2 (analysed %).
- Air and O? supply pressures.
- Plateau Pressure.
- Resistance and compliance.
- Accuracy pressure readings equal or better than: ± 2 cm H?O.
- Accuracy volume readings equal or better than: ± 10% or 1 ml, whichever is greater.
- Accuracy O? concentration readings equal or better than: ± 3%.
- Minimum of 24 hours trends storage.
- Event log which stores past 100 alarm events or events over the past 24 hrs.
- Display of compliance, resistance waveforms: Pressure – Time, Flow – Time, Volume – Time.
- Display of time loops: Pressure – Volume, Flow – Volume, Pressure – Flow.
AUDIO AND VISUAL ALARMS
The unit supports alarms for the below listed parameters and features:
- Audio-visual alarms for all measured and monitored parameters.
- Disabled alarms should be indicated on the display.
- Silenced alarms should reactivate automatically within two minutes.
- The volume of the alarm cannot be turned down to an inaudible level.
- Airway pressure (adjustable, high, low, peak pressure).
- Respiratory rate (adjustable, high).
- Tidal Volume (adjustable, high/low).
- Minute Volume (adjustable, high/low).
- FiO2 (adjustable, high/low).
- PEEP (adjustable, high/low).
- Apnoea.
- Inverse IE ratio.
- Patient circuit disconnection.
- Occlusion and/or leakage in the breathing circuit.
- Sensor failure.
- General failure to operate.
- Failure in O? and/or air supply.
- Low, empty or battery failure.
- AC power supply failure.
ESTIMATED LIFE SPAN
More than seven years.
WARRANTY
Two years from shipping date.
ENVIRONMENTAL CONDITIONS
- Operating conditions: 10°C - 40 °C / 0% – 95% RH
- Storage conditions: -25°C - 70°C / 0% – 95% RH
- Automatic compensation up to 5,000 meter above sea level.
- Ingress protection rating: IP21
WEIGHT AND VOLUME
Weight: 85 kg.
Volume: 1,000.00 dm³.