Anaesthesia machine,closed circuit,w/acc

Anaesthesia machine,closed circuit,w/acc

Product Code : SCL-MAI-12232
GENERAL DESCRIPTION
Anaesthesia machine closed or semi-closed circuit configuration.

INTENDED USE
Anaesthesia units dispense a mixture of gases and vapours and vary the proportions to control a patient's level of consciousness and/or analgesia during surgical procedures. Anaesthesia units primarily perform the following four functions:
- Provide oxygen (O?) to the patient
- Blend gas mixtures, in addition to O?, that include air or nitrous oxide (N?O) along with an anaesthetic vapor.
- Facilitate spontaneous, controlled, or assisted ventilation while using these gas mixtures.
- Reduce, if not eliminate, anaesthesia-related risks to the patient and clinical staff.
The patient is anesthetized by inspiring a mixture of O?, the vapor of a volatile liquid halogenated hydrocarbon anaesthetic, and, if necessary, N?O and other gases. Because normal breathing is routinely depressed by anaesthetic agents and by muscle relaxants administered in conjunction with them, respiratory assistance is also provided via either an automatic ventilator or by manual compression of the reservoir bag.

TECHNICAL SPECIFICATIONS
Suitable for adult, paediatric and newborn patients.
The unit is mounted on a trolley with minimum of four (4) anti-static swivel castors.
A minimum of two of the castors are provided with breaks.
The unit is equipped with an upper shelf.
The cart has been proved with handles for manoeuvring.
A side rail for mounting accessories is included.
The unit has an adjustable patient-circuit support arm.
The unit is provided with a minimum of three (3) gas inlets (O?, N?O and Air).
Gas inlet connections are compliant with DISS or NIST (needs to be specified when ordering).
The unit is provided with a minimum of two (2) gas cylinder yokes (O?, and N?O).
Holders for bottles are to include a secure locking mechanism.
The minimum gas supply input pressure range should be between 2.8 – 6 bar (41 - 87 psi).
The unit are equipped with gas supply gauges with scales allowing easy readout.
All pipeline connections have diameter-indexed safety systems (DISSs), or another means of preventing connection of dangerous gases.
The unit is equipped with a minimum of four (4) analogue rotameters, two (2) for O?, one (1) for air and one (1) for N?O with programmed parameters visualization.
The O? and N?O flowmeters have a minimum range of 0.0 – 10 L/min, and a resolution of at least 0.2 L/min.
Externally supplied N?O, O?, and air mixture ratios are fully controllable. There are separate controls for the flow of: O?, N?0, and anaesthetic agents.
Flows and the mixture ratios determined from flowmeter settings are accurate to within ± 10% of set values or better.
The unit has the ability to carry out self-diagnosis and integrity testing, including a compliance and leakage test.
The unit is equipped with a minimum of two (2) vaporizer slots with a Selectatec mounting system.
The vaporizers are provided with an interlock system, preventing the use of more than one vaporizer simultaneously.
One (1) Halothane vaporizer is included with the unit.
The unit basic is equipped with a passive scavenging system and optional allow for an active scavenging system.
The unit is equipped with a soda lime CO? absorber.
The unit is equipped with a non-return and a three-way valve, including the connection tube.
A built-in rechargeable battery allows for an autonomy of at least one (1) hour.
The unit will switch back and forth between battery and mains operation in case of power failures. Recharging resumes automatically when extern power becomes available.
The unit is designed and constructed in such a way that it is able to withstand frequent disinfection with hospital-grade products.
Fuses are fitted on both live and neutral supply lines.
Power requirements: 100 - 240 Volts - 50/60 Hz (not necessarily in a single unit).

VENTILATOR
The unit should be equipped with a ventilator suitable for adult, paediatric, and newborn patients and minimally supporting the following ventilation modes:
- Volume Controlled Mode.
- Stand-by Mode.
- Manual Mode.
The unit should be able to provide a minimum guaranteed tidal volume of 20 – 1,500 mL.
The unit should be able to support a minimum ventilation frequency range between 1 – 100 bpm.
The unit supports an I/E ratio 4:1 to 1:10.
The unit supports an inspiration pressure of 5 – 70 cmH?O.
The unit supports a peak inspiratory flow range of 1 – 70 L/min.
The unit supports PEEP with a range of 3 – 30 cmH?O.
The unit supports pressure triggers in the range of 0 – 20 cmH?O.
Pressure gauges have a range of 0 – 100 cmH?O and an accuracy of: ±2.5 % or better.

MONITORED PARAMETERS
The ventilation monitors following parameters, through a combination of numerical and waveforms:
- Respiratory rate (spontaneous and mechanical).
- Tidal volume (inspired and expired).
- Minute volume (spontaneous and mechanical).
- Airway pressure.
- PEEP.
- Compliance.
- O? Concentration.
- I:E ratio.
- Inspiration and expiration times.
- End-tidal CO? (capnography).
- FiO? dynamic.
- Plateau pressure.
- Trend display facility for at least the last 24 hours, with minimum 5 minutes resolution.
- Automatic compliance and leakage compensation for circuit and tubes.

ALARM AND SAFETY FUNCTIONALITIES
Alarms are categorized in three categories: caution, advisory and alarm.
The unit is equipped with audio and visual alarms for the following parameters:
- High airway pressure alarm.
- Sub atmospheric pressure alarm.
- Respiratory rate alarm.
- Minute volume alarm.
- Tidal volume alarm.
- Expiratory flow alarm.
- FiO? supply failure alarm.
- O? supply failure alarm.
- Low pressure/apnoea alarm.
- External O? gas supply failure alarm.
- Low battery alarm.
- Power failure alarm.
- Sensor disconnected alarm.
- System or sensor failures alarm.
The unit has a provision that prevents an anaesthesia machine from being set to dispense a hypoxic mixture. The N?O and O? flow controls are interlocked so that the proportion of O? to N?O can never fall below a minimum value of 25% to produce a hypoxic breathing mixture.
The unit has a provision that protects the patient from inadequate O? supply. If the O? supply pressure drops below 1.7 - 2.1 Bar (25 to 30 psi), the unit decreases or shuts off the flow of other gases and activates an alarm.
The unit is equipped with an Adjustable Pressure Limiting (APL) valve to prevent delivery from too high-pressured gas.
The unit should have a provision for emergency O? by-pass.

DISPLAYED PARAMETERS
The unit is equipped with a flat panel colour display indicating the following parameters and traces:
Three (3) traces against time: pressure, volume, and flow.
Two (2) two-axis displays: Pressure-Volume and Flow-Volume.
Values for all monitored parameters.
Alarm settings for all monitored parameters.
Alarms or errors in ventilation or anaesthesia parameters.
Current time.
Ventilator mode.
Battery status.
System events.

SUPPLIED WITH
Instructions for assembly, use and maintenance in English.
1 x Plastic protective dustcover.
1 x newborn reusable breathing circuit (tubes/balloon/valves/mask).
1 x paediatric reusable breathing circuit (tubes/balloon/valves/mask).
1 x adult reusable breathing circuit (tubes/balloon/valves/mask).
2 x soda lime absorber grains, pack/2.5 kg (depending on model supplied).
1 x spare parts/maintenance kit (air filters, tubing, O-rings).
1 x set of spare fuses, if required.



OPTIONAL ACCESSORIES AVAILABLE
At extra costs additional optional features can be requested:
- A fully integrated compressor. (depending on the model supplied).
- A fully integrated active scavenging system.
- A fully integrated SpO? system.
- A fully integrated gas measuring system.
- Integrated end-tidal CO? with capnography.

ESTIMATED LIFE SPAN
8 – 10 years.

WARRANTY
Two years from shipping date.

ENVIRONMENTAL CONDITIONS
- Operating conditions: 15°C - 35°C / 95% RH.
- Storage conditions: -5°C - 40°C / 95% RH.
- Atmospheric pressure: 500 ~ 1060 hPa.

WEIGHT AND VOLUME
Weight: 150 kg.
Volume: 1,055.00 dm³.

ESTIMATED DELIVERY LEAD TIME
120 days

   
Supply of Anaesthesia machine,closed circuit,w/acc from to Finland, France, Tanzania, Thailand, Sri Lanka, Congo, Costa Rica, Ivory Coast, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominica, Afghanistan, Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, United Arab Emirates (UAE, Dubai), United Kingdom (London), United States, Uruguay, Uzbekistan, Vanuatu, Belarus, Belgium, Belize, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burma Myanmar, East Timor, Ecuador, Egypt, El Salvador, Equatorial Guinea, Eritrea, Estonia, Ethiopia (Addis Ababa), Fiji, Burundi, Cambodia, Cameroon, Canada, Cape Verde, Central African Republic, Chad, Chile, Colombia. Math Teaching Kit suppliers in Luxembourg, Macedonia, Madagascar, Malawi (Lilongwe), Malaysia (Kuala Lumpur), Maldives, Mali, Malta, Marshall Islands, Mauritania, Mexico, Micronesia, Moldova, Monaco, Mongolia, Montenegro, Morocco, Mozambique, Namibia, Comoros, Congo. Math Teaching Kit manufacturers in Dominican Republic, Gabon, Gambia, Georgia, Germany, Ghana, Greece, Grenada, Guatemala, Guinea, Guinea-Bissau, Guyana, Haiti, Honduras, Hungary, Iceland, Indonesia, Iran, Iraq, Niger, Nigeria (Abuja), Norway, Oman, Palau, Panama, Papua New Guinea, Paraguay, Peru, Philippines (Manila), Poland, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya (Nairobi), Kiribati, Korea, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon, Lesotho, Liberia, Liechtenstein, Lithuania, Mauritius, Nauru, Nepal, Netherlands, New Zealand, Nicaragua, Portugal, Qatar, Romania, Russia, Rwanda (Kigali). Math Teaching Kit exportets to Saint Kitts and Nevis, Saint Lucia, Saint Vincent and the Grenadines, Samoa, San Marino, Sao Tome and Principe, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia, Slovenia, Solomon Islands, Somalia, South Africa, South Sudan, Spain, Sudan, Suriname, Swaziland, Sweden, Switzerland, Syria, Tajikistan, Togo, Tonga, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Tuvalu, Uganda (Kampala), Ukraine, Venezuela, Vietnam, Yemen , Zambia (Lusaka), Zimbabwe