Electrosurgical unit,w/access
.png)
Product Code : SCL-MAI-12225
GENERAL DESCRIPTION
Electrosurgical unit, for general surgery including a trolley.
INTENDED USE
Electrosurgical units are used for surgical cutting and for controlling bleeding by causing coagulation (haemostasis) at the surgical site. They deliver high-frequency electrical current through an active electrode tip, causing vaporization, desiccation, or charring by resistive heating in the target tissue. Electrosurgery is useful for nearly all surgical procedures, both open and laparoscopic. The haemostatic effect of electrosurgery makes it useful for procedures on organs with oozing capillary beds, such as the liver, spleen, thyroid, and lungs, as well as for open-heart procedures in which large amounts of anticoagulants are used.
TECHNICAL SPECIFICATIONS
Solid-state/microprocessor-controlled frequency generator.
Monopolar and bipolar outputs, electrically isolated from ground.
Minimum output frequency is higher than 400kHz.
Monopolar modes include pure cut, blend, and coagulate (soft, contact and spray).
Bipolar mode includes coagulate and cut mode.
Maximum monopolar cut power output maximum 300 W.
Maximum monopolar coagulation power output maximum 100 W.
Maximum bipolar power output maximum 100 W.
Hand switch mode when button-activated probes are connected.
Foot switch that can operate in monopolar and bipolar modes.
Yellow buttons/pedals for cut and blue buttons/pedals for coagulate.
Grounding pad/return electrode monitored for patient connection.
Front panel allows mode selection, power settings and on/off.
Display shows output power, system errors and electrode failure.
Automatic shut off/generator deactivation on grounding pad/return electrode connection failure.
Audible and visual indicators of activation and alarms.
Self-test mode.
Protection against defibrillator discharges.
Power requirements: 100 - 240 Volts - 50/60 Hz (indicate exact requirements at the time of ordering).
ESTIMATED LIFE SPAN
5 years.
WARRANTY
Depending on the model one or two years from shipping date.
ENVIRONMENTAL CONDITIONS
- Operating conditions: 10°C - 40°C / 30% – 75% RH.
- Storage conditions: -10°C - 60°C / 30% – 90% RH.
- Atmospheric pressure: 700 ~ 7060 hPa.
WEIGHT AND VOLUME
Weight: 28 kg.
Volume: 160.00 dm³.
ESTIMATED DELIVERY LEAD TIME
120 days.