Product Code : SCL-MAI-12412 GENERAL DESCRIPTION Hand held, battery powered device that measures the oxygen concentration in a flow of gas from a medical gas source or, with adapters, through a medical gas-flow device such as a ventilator or anaesthesia system. The intended application is checking gas supply equipment performance.
TECHNICAL SPECIFICATIONS Operational Characteristics: Ultrasonic oxygen sensing technology, with permanent sensing chamber. O2 measurement to include the range 21 - 96%. O2 resolution: 0.1%. O2 accuracy: within ± 1.5% of full scale. Flow rate measurement: up to 10 LPM Flow rate accuracy within ± 0.2 LPM Flow rate resolution 0.1 LPM Suitable for measuring gas supply with pressure up to 50 psi (344 kPa) Response time < 20 s. Warm-up time < 1 s. Display visualizing O2 concentration, flow rate, system messages and battery status.
Electrical characteristics: Operated by 2 x AA alkaline replaceable single use batteries. Battery life > 1000 hours continuous use. Automatic power-off when not in use.
Casing and environment IPX1 ingress protection. Suitable for cleaning and disinfection with hospital-grade cleaning products.
SUPPLIED WITH 1 x protective case Connector tube for gas sampling 1 x set single use batteries (2 AA batteries) 1 x Instructions for Use, hard and soft copies in English, French, Spanish, German, Dutch and Italian. 1 x Contact details of manufacturer.
ENVIRONMENTAL CONDITIONS Capable of being stored continuously in ambient temperature from -15°C to 60°C, RH: 0 to 95% Capable of operating continuously in ambient temperature from 15°C to 40°C, RH: 0 to 95%